Patient monitor BPM-9520-Patient Monitor-Shenzhen Bestman Instrument Co.,ltd
 Location: Home >
  • 9520-1.jpg
Patient monitor BPM-9520
Description

BPM-9520 Multi-Parameter Patient Monitor

 

8.4''high resolution color TFT display 

Light weight, user-friendly design, easy operation mode

Pacemaker detection, arrhythmia, S-T segment analysis 

Digital SpO2 technology with better   

performances of anti-electrosurgical  interference, anti-motion and anti-low infusion 

Calculation function of drug concentration, drug titration and hemodynamic status   

Wired or wireless network   

Intelligent audible and visible alarm , events recallable   

96 hours trend graphic recall, 96 hours trend table recall   

For adult, pediatric and neonate

 

Standard configuration:  ECG,RESP,NIBP,SPO2,PR,TEMP

Optional function:  IBP,EtCO2,Cardiac output,printer

Specification:
Dimension:                   250(H)X146(W)X240(D)mm
Weight:                         4Kgs (with battery)


ECG
Measuring range:         Adult  0~300bpm
                       Neonate and pediatric   0~350bpm
Band width:                   0.05Hz~100Hz 
Accuracy:                     ±1%
Lead Type:                    5-lead and 3-lead selectable
Lead selection:         standard I,II,III; augmented limb lead 
                            aVR, aVL, aVF;Chest lead V1~V6
Resolution:                   1 bpm
Trend monitoring time:no less than 96 hours
ST-Segment analysis:    -2.0 mV ~2.0mV; auto 
Arrhythmia analysis:    13 kinds of Arrhythmia analysis
Scanning speed:        12.5, 25, 50mm/Second
Gain selection:         x1/2, x1, x2; auto 
Protection:            Withstand 4000 VAC/ 50Hz voltage 
                     Voltage in isolation against
                     Defibrillation and electrosurgical
                     interference
ECG cable :           adults, pediatric and neonate 
Alarm Range:         15-350 bpm
ARR Alarm:          Audible and visible alarm , 
                    events recallable


NIBP
Method:                 Automatic Oscillometric
Operation Model:          manual, automatic, continuous  
Auto measurement 
time interval:             Adjustable 1 /2 /3/4
                                /5/10/15/30/60/90/120/240/480 Min.
Measuring Unit:          mmHg /kPa selectable
Measurement types:       Systolic, Diastolic, Mean
Measuring Range:         Adult:    10~270mmHg
                                     Pediatric:10~200mmHg
                                     Neonate:10~135mmHg
Cuff pressure range :      0~320mmHg
Cuff type:                     Adult cuff
                                     Pediatric
                                     Neonate 
Cuff inflation time:       < 10 seconds 
Cuff deflation time:       <20 seconds  
Over-Pressure 
Protection:              Yes 
Alarm:                 Upper alarm,lower alarm,Auto-memory


TEMP
Measuring range:         0~ 50 ℃ ( 32~122℉  )
Probe Type:                 Body temperature probe (standard)
                                     Esophageal temperature probe (Optional)
                                     Rectal Temperature probe  (Optional)
Resolution:              0.1℃
Accuracy:               ±0.1℃ 
Channel:                Dual-channel,Provide T1,T2
Alarm Range:            0~50℃


SpO2
Measuring range          0~100% 
Resolution:              1% 
Accuracy:               ±1% ( 70%~100%) 
                       ±2% (0%~69% )
PR measuring range:      0~255 bpm
PR accuracy:            ±1 bpm
Sensor Type:            Adult  sensor (standard)
                             Pediatric  sensor(optional) 
                             Neonate  sensor(optional) 
Alarm Range:           0~100%


RESP
Display:               waveform display and numeric display
Operation Mode:        Auto/Manual
Method:               Trans-thoracic impedance 
RR Measuring range:    0~150bpm
Accuracy:             ±1 bpm
Apnea alarm 
threshold:             10s,15s,20s(default),25s,30s,40s
Band Width:           0.2-2.5Hz (-3dB)
Sweep speed:          6.25mm/s,12.5mm/s,25mm/s,50mm/s
Alarm Range:         Adult:6-120rpm
Pediatric and Neonate:   6-150rpm


IBP (Optional)
Measurement range:    - 10 ~ 300 mmHg
Channel:              2 channels
Sensitivity:            5 μ V/V/mmHg
Impedance range       300~3000Ω
Accuracy:             ±1 mmHg or ±2%
                     (exclusive of transducer)
Alarm Range:         -10 ~ 300 mmHg


EtCO2 (Optional)
Measurement range:    0 ~ 150 mmHg
Accuracy:            1rpm
Sample flowerate:     50 ml/min
Sample rate accuracy   ±15% or 15 ml/min.
Respiration rate:       0 ~ 120 rpm
Respiration accuracy:  ±2 rpm (0 ~70 rpm)
                   ±5 rpm (>70 rpm)
Respiration time:     < 60 ms