Patient monitor BPM-9010-Patient Monitor-Shenzhen Bestman Instrument Co.,ltd
 Location: Home >
  • 9010.jpg
Patient monitor BPM-9010
Description

BPM-9010  Multi-Parameter Patient Monitor


 12.1''high resolution color TFT display 
 96 hours trend graphic data, 12-lead ECG analysis
 Arrhythmia and S-T segment analysis 
 Anti-defibrillation, Anti-high frequency electrotome 
 Multi-language,easy operate,durable and stable 
 Calculation function of drug concentration , drug titration and hemodynamic status    
 Wired or wireless network    
 Three-level Audible and visible alarm,events recallable , Nurse Call function
 AC/DC power supply, built-in high capacity battery


Standard configuration:  ECG,RESP,NIBP,SPO2,PR,TEMP
Optional function:  IBP,EtCO2,Cardiac output,printer

Dimension:                   272(H) X 327(W) X 145(D) (mm)

Weight:                         6.5Kgs (with battery)

 

ECG

Measuring range:              Adult   0~300bpm     

Neonate and pediatric   0~350bpm

Band width:                  0.05Hz~100Hz

Accuracy:                     ±1%

Lead Type:                   5-lead and 3-lead selectable

Lead selection:             standard III,III; augmented limb lead

                         aVR, aVL, aVFChest lead V1~V6

Resolution:               1 bpm

Trend monitoring time:no less than 96 hours

ST-Segment analysis:        -2.0 mV ~2.0mV; auto

Arrhythmia analysis:        13 kinds of Arrhythmia analysis

Scanning speed:            12.5, 25,50mm/Second

Gain selection:             x1/2, x1, x2; auto

Protection:                Withstand 4000 VAC/ 50Hz voltage

                                     Voltage in isolation against

                                     Defibrillation and electrosurgical

                                     interference

ECG cable :                adults, pediatric and neonate

Alarm Range:               15-350 bpm

ARR Alarm:                Audible and visible alarm,

                                     events recallable

 

NIBP

Method:                    Automatic Oscillometric

Operation Model:             manual, automatic, continuous 

Auto measurement

time interval:                Adjustable1 /2 /3/4

                                     /5/10/15/30/60/90/120/240/480 Min.

Measuring Unit:              mmHg /kPa selectable

Measurement types:           Systolic, Diastolic, Mean

Measuring Range:            Adult:     10~270mmHg

                                     Pediatric:10~200mmHg

                                     Neonate: 10~135mmHg

Cuff pressure range :          0~320mmHg

Cuff type:                   Adult cuff (standard)

                                     Pediatric and Neonate(Optional)

                                     Pediatric and Neonate(Optional)

Cuff inflation time:           < 10 seconds

Cuff deflation time:           <20 seconds 

Over-Pressure

Protection:                  Yes

Alarm:                     Upper alarm,lower alarm,Auto-memory

 

TEMP

Measuring range:            0~50 ( 32~122  )

Probe Type               Body temperature probe (standard)

                                     Esophageal temperature probe(Optional)

                                    Rectal Temperature probe (Optional)

Resolution:                  0.1

Accuracy:                   ±0.1

Channel:                    Dual-channe,Provide T1, T2

Alarm Range:                0~50

 

SpO2

Measuring range              0~100%

Resolution:                  1%

Accuracy:                   ±1% ( 70%~100%)

                           ±2% (0%~69% )

PR measuring range:          0~255 bpm

PR accuracy:                ±1 bpm

Sensor Type:                Adult  sensor (standard)

                                     Pediatric  sensor(optional)

                                     Neonate  sensor(optional)

Alarm Range:                0~100%

 

RESP

Display:                    waveform display and numeric display

Operation Mode:             Auto/Manual

Method:                    Trans-thoracic impedance

RR Measuring range:          0~150bpm

Accuracy:                    ±1 bpm

Apnea alarm threshold: 10s,15s,20s(default),25s,30s,40s

Band Width:                 0.2-2.5Hz (-3dB)

Sweep speed:                6.25mm/s,12.5mm/s,25mm/s,50mm/s

Alarm Range:                Adult:6-120rpm

                           Pediatric and Neonate:6-150rpm

 

IBP (Optional)

Measurement range:           - 10 ~ 300 mmHg

Channel:                    2 channels

Sensitivity:                  5 μ V/V/mmHg

Impedance range             300~3000Ω

Accuracy:                   ±1 mmHg or ±2%

                           (exclusive of transducer)

Alarm Range:                -10 ~ 300 mmHg

 

EtCO2 (Optional)

Measurement range:          0 ~ 150 mmHg

Accuracy:                  1rpm

Sample flowerate:            50 ml/min

Sample rate accuracy         ±15% or 15 ml/min.

Respiration rate:             0 ~ 120 rpm

Respiration accuracy:         ±2 rpm (0 ~70 rpm)

                          ±5 rpm (70 rpm)

Respiration time:             60 ms

Delay time:                2s